With the CL541-PLUS you can check and calibrate all your thermocouple instruments and measure thermocouple sensors. Automatic indication of connections on the display for simple hookups. Take it without worry into the shop, plant or field. It comes protected with a rubber boot and rugged, low profile switch. Easy to operate even in the dark areas of the plant with the backlit display. Stop carrying around a millivolt source and thermocouple tables, you can calibrate directly in temperature (°C and °F).

The CL541-PLUS works with the T/Cs you use including types J, K, T and E. Easily set any value quickly to within 0.1° with the adjustable digital potentiometer "DIAL" plus store any three temperatures for instant recall. Calibrate quickly with automatic output stepping. Choose between 2, 3, 5, 11 and 21 steps to automatically increment the output in 100%, 50%, 25%, 10% or 5% of span. Select the step time to match your system from 5, 6, 7, 8, 10, 15, 20, 25, 30 and 60 seconds. Connect directly to the thermocouple inputs of smart transmitters, PLCs, controllers and multichannel recorders and verify their outputs or displays. Measure and trouble shoot thermocouple sensor connections and find broken wires or high resistance junctions. Connect your sensor and the CL541-PLUS automatically measures the thermocouple in degrees C or F.

SPECIFICATIONS

(Unless otherwise indicated all specifications are rated from a nominal 3°C, 70% RH)

Accuracy: ±(0.008% of Reading + 0.006 mV) Cold Junction

Compensation: ±0.1°F (±0.06°C)

Millivolt Range: -13.000 to 80.000 mV Operating Temperature

Range: -25 to 60°C (-10 to 140°F)

Relative Humidity Range: 10% ≤RH ≤90% (0 to 35°C), non-condensing; 10% ≤RH ≤70% (35 to 60°C), non-condensing

Dimensions: 143 L x 76.2 W x 40.64 mm H (5.63 x 3.00 x 1.60")

Weight: 12.1 oz (including boot and batteries)

Power: 4 "AA" Alkaline 1.5V (included)

Battery Life: 50 Hours

Low Battery: Low battery indication with nominal 1 hour of operation left

Protection Against Misconnection: Over-voltage protection to 60 Vdc (rated for 30 seconds)

Display: High contrast graphic liquid crystal display. LED backlighting for use in low lit areas.

Read

Input Impedance: >10 Megohms

Open Thermocouple

    Threshold: 10,000Ω nominal

    Pulse: <10 microamp pulse for 300 milliseconds

Normal Mode Rejection: 50/60 Hz, 50 dB

Common Mode Rejection: 50/60 Hz, 120 dB Source

Output Impedance: < 0.3Ω

Source Current: > 20 mA (drives 80 mV into 10Ω)

Noise: ≤4 microvolts p-p for frequencies of 10 Hz or below